Description
Conductivity, TDS, Temperature In-line Analyzer. 1/4 DIN, IP65, panel mount TDS/Temp analyzer with two sets of 2 wire 4-20mA current output, 2 sets of programmable on/off relay, RS485 output and backlight.
- Graphic LCD with manual backlight displays conductivity/TDS,
temperature, relay status and transmitter current mA output simultaneously. - Convenient one point calibration and calibration data is stored in memory and is ready for use on power up.
- Two sets of Isolated and reversible 4-20 mA current output for conductivity and temperature allows it to be used as a proportional output.
- 2 programmable on/off relays conductivity.
- RS-485 output for remote access allows multiple units to be connected and ganged together.
- Password protection.
Specifications
Conductivity Range K=0.01-1 cm = 0.000 to 300.9 uS/cm / 0.000 to 150.0 ppm (over 2 ranges)
Conductivity Resolution K=0.01-1 cm = 0.001 to 0.1 uS/cm / 0.001 to 0.1 ppm (over 2 ranges)
Conductivity Accuracy K=0.01-1 cm = ±0.5% FS ± 1 digit
Conductivity Range K=0.1-1 cm = 10.00 uS/cm to 3.000 mS/cm / 5.00 ppm to 1.500 ppt (over 2 ranges)
Conductivity Resolution K=0.1-1 cm = 0.01 uS/cm to 0.001 mS/cm / 0.001 ppm 0.1 ppt (over 2 ranges)
Conductivity Accuracy K=0.1-1 cm = ± 0.5% FS ± 1 digit
Conductivity Range K=1.0-1 cm = 100 uS/cm to 30.00 mS/cm / 50 ppm to 15.00 ppt (over 2 ranges)
Conductivity Resolution K=1.0-1 cm = 0.1 uS/cm to 0.01 mS/cm / 0.1 ppm to 0.01 ppt (over 2 ranges)
Conductivity Accuracy K=1.0-1 cm = ±0.5% FS ± 1 digit
Conductivity Range K=10-1 cm = 1.0 to 300 mS/cm / 0.000 to 150.0 ppt (over 2 ranges)
Conductivity Resolution K=10-1 cm = 0.001 to 0.1 mS/cm / 0.001 to 0.1 ppt(over 2 ranges)
Conductivity Accuracy K=10-1 cm = ±0.5% FS ± 1 digit
Temperature Range = -10.0 to 120.0 °C
Temperature Resolution = 0.1 °C
Temperature Accuracy = ±0.3 °C
Temperature Compensation Auto = -10.0 to 120.0°C (Thermistor, 10K ohm at 25.0 °C)
Cell Constant = 0.01, 0.1 , 1.0 ,10cm(exp)-1, 2 wire or 4 wire
Reference Termperature = From 10.0 to 29.0 °C,user selectable
Temperature Coefficient = From 10.0 to 29 °C, user selectable
Output = Two current outputs for conducitivity and temperature
Current Output Range = 4 to 20 mA(isolated),programmable span
Current Output Scale = User programmable ,linear
Maximum Load = 500 ohm /500V DC
Control Type = 2 ON/OFF controls
Relay Output = 5A at 115V AC or 2.5A at 220V AC, resistive load only
Display = 128 x 64 graphical LCD with backlight
Security = 4-dight password
Power = 115 V AC or 230V AC at 50/60 Hz
Ambient Temperature Range = From 0 to 50 °C
Weight = 950 grams
Case = IP 65 ,1/4 DIN case, 148mm depth
Documentation and Supports
6312CTB.pdf